EasyManua.ls Logo

Fluke 433 - Page 16

Fluke 433
74 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Fluke 433, 434, 435
Service Manual
2-8
Harmonics Measurement Range Resolution Accuracy
Harmonic order (n) DC, 1..50 Grouping: Harmonic groups according to IEC 61000-4-7
Inter-Harmonic order Off, 1..49 Grouping: Harmonic and Interharmonic subgroups according to IEC
61000-4-7
Vrms Relative (%f):
Fluke 435 Absolute:
Fluke 433/434 Absolute:
0.0 … 100.0%
0.0 … 1000 Vrms
0.0 … 1000 Vrms
0.1%
0.1 Vrms
0.1 Vrms
± 0.1% ± n x 0.1%
(± 0.4% for %r)
± 0.05% of nominal
voltage if < 1% of
nominal voltage
± 5% if 1% of nominal
voltage
± 5% ± 2 counts
Arms Relative (%f):
Absolute:
0.0 … 100.0%
0.0 … 4000 mV x clamp
scaling
0.1%
1 mVrms x clamp scaling
± 0.1% ± n x 0.1%
(± 0.4% for %r)
± 5% ± 5 counts
Watts Relative:
(Harmonics only)
Watts Absolute:
(Harmonics only)
0.0 … 100.0%
depends on clamp and
voltage scaling
0.1%
± n x 2%
± 5% ± n x 2% ± 10
counts
DC Relative:
Fluke 435 Absolute V:
Fluke 433/434 Absolute
V:
Absolute A:
Absolute W:
0.0 … 100.0%
0.0 … 1000V
0.0 … 1000V
0.0 … 4000 mV x clamp
scaling
depends on clamp and
voltage scaling
0.1%
0.1V
0.1V
1 mVrms x clamp scaling
0.1V
depends on scaling
± 0.1% V and A (± 2%
Watt)
± 0.2% of nominal
voltage
± 5% ± 10 counts
± 5% ± 10 counts
± 5% ± 10 counts
THD
(n=40)
(relative %f or
%r)
0.0 … 100.0 % 0.1% ± 2.5% V and A (± 5%
Watt)
Hz 0 … 3500 Hz 1 Hz ± 1Hz
Phase angle
Fluke 435
Fluke 433/434
-360º ... +0º
-360º ... +0º
± n × 1º (
8
)
± n × 1.5º (
8
)

Related product manuals