EasyManua.ls Logo

Fluke 433 - Page 36

Fluke 433
74 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Fluke 433, 434, 435
Service Manual
3-12
13. Fluke 433/434: check for a voltage readout V rms between 0.0 ... 1.2 V.
Fluke 435: check for a voltage readout V rms between 0.00 ... 0.23 V.
14. Check the 230 V range according to the table below.
Table 3-7. Accuracy Check of Voltage Channel N (Neutral)
Set Calibrator to Readout at Voltage Channel N
0 Hz, 0 V, OPR Fluke 433/434: 0.0 ... 1.2 V, Fluke 435: 0.00 ... 0.23 V
115 V, 50 Hz, OPR Fluke 433/434: 113.8 ... 116.2 V, Fluke 435: 114.77 ... 115.23 V
230 V, 50 Hz, OPR Fluke 433/434: 228.8 ... 231.2 V, 49.99 ... 50.01 Hz
Fluke 435: 229.77 ... 230.23 V, 49.990 ... 50.010 Hz
460 V, 50 Hz, OPR Fluke 433/434: 458.8 ... 461.2 V, 49.99 ... 50.01 Hz
Fluke 435: 459.77 ... 460.23 V, 49.990 ... 50.010 Hz
15. Set the Calibrator to STBY.
Optional Test. Bandwidth Check of Channel N (*):
Note: for Fluke 435 you must use the Fluke 5700A Calibrator. This instrument is able to
generate 219.99999 V as a maximum at 3 kHz. Therefor a 50 Hz reference voltage of 215
V must be adjusted.
16. Set the Calibrator to 230 V (5700A: 215 V), 50 Hz and OPR.
17. Now check the voltage channel N. Use the Calibrator’s field edit function (5500A:
FIELD EDIT key, 5700A: AMPL/FREQ key) to adjust the Calibrator to an Analyzer
readout of 230.0 V / 215.00 V.
18. Increase the frequency to 3 kHz.
Fluke 433/434: check for a readout of 218.3 V or more.
Fluke 435: check for a readout of 204.25 V or more.
19. Set the Calibrator to STBY.
3.7.4 Verification of voltage inputs in 400 V range
Accuracy Check of Channel A/L1, B/L2, C/L3:
1. Set the Analyzer to 400 V, 50 Hz nominal (SETUP menu, arrow and ENTER keys).
Press MENU, select Volts/Amps/Hertz, press F5 - OK.
2. To check the A/L1, B/L2, C/L3 inputs: connect the N input to Ground (See Figure 3-
3).
3. Set the Calibrator to 0 Hz, 0 V and then to OPR.
4. Fluke 433/434: check for a voltage readout V rms between 0.0 ... 2.0 V.
Fluke 435: check for a voltage readout V rms between 0.00 ... 0.40 V.
5. Check the 400 V range according to the table below.

Related product manuals