Fluke 433, 434, 435
Service Manual
4-6
Cal Name
Name of the selected calibration step, e.g. Warming Up
(CL nnnn) Number of the calibration step
Status (...) can be:
IDLE (valid) After (re)entering this step, the calibration process is not started.
The calibration data of this step are valid. This means that the
last time this step was done, the calibration was successful. It
does not necessarily mean that the unit meets the specifications
related to this step!
IDLE (invalid) After (re)entering this step, the calibration process is not started.
The calibration data are invalid. This means that the last time this
step was done, the calibration was not successful. Most probably
the unit will not meet the specifications if the actual calibration
data are saved.
BUSY aaa% bbb% Calibration adjustment step in progress; progress % for Input A
and Input B. During Warming Up the elapsed time is shown.
READY Calibration adjustment step finished.
Functions of the keys F1-F4 are:
F1
PREV select the previous step
F2
NEXT select the next step
F3
CAL start the calibration adjustment of the actual step
F5
EXIT leave the calibration mode