EasyManua.ls Logo

Fluke 51 - Electrical; Environmental; General Specifications

Fluke 51
28 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
51, 52, 53, 54 Series II
Service Manual
4
Specifications
General Specifications
Weight 280 g (10 oz)
Dimensions
(without holster)
2.8 cm × 7.8 cm × 16.2 cm
(1.1 in × 3 in × 6.4 in)
Battery 3 AA batteries
Certification P,
)
(TÜV pending)
Safety CSA C22.2 No. 1010.1 1992
EN 61010 Amendments 1, 2
CAT I OVERVOLTAGE (Installation) CATEGORY I,
Pollution Degree 2 per IEC1010-1*
* Refers to the level of Impulse Withstand Voltage protection provided. Equipment of OVERVOLTAGE
CATEGORY I is equipment for connection to circuits in which measures are taken to limit the transient
over voltages to an appropriate low level.
Environmental
Operating Temperature 10
o
C to +50
o
C (+14
o
F to +122
o
F)
Storage Temperature 40
o
C to +60
o
C (40
o
F to +140
o
F)
Humidity
Non Condensing (< +10
o
C) (< +50
o
F)
95% RH (+10
o
C to +30
o
C) (+50
o
F to +86
o
F)
75% RH (+30
o
C to +40
o
C) (+86
o
F to +104
o
F)
45% RH (+40
o
C to +50
o
C) (+104
o
F to +122
o
F)
(Without Condensation)
Electrical
Measurement
Range
J-type: 210
o
C to +1200
o
C (346
o
F to +2192
o
F)
K-type: 200
o
C to +1372
o
C (328
o
F to +2501
o
F)
T-type: 250
o
C to +400
o
C (418
o
F to +752
o
F)
E-type: 150
o
C to +1000
o
C (238
o
F to +1832
o
F)
N-type: 200
o
C to +1300
o
C (328
o
F to +2372
o
F)
R- and S-type: 0
o
C to +1767
o
C (+32
o
F to +3212
o
F)
Display
Resolution
0.1
o
C /
o
F / K < 1000
o
1.0
o
C /
o
F / K 1000
o
Measurement
Accuracy
J-, K-, T-, E-, and N-type: ±[0.05 % of reading + 0.3
o
C (0.5
o
F)]
[below 100
o
C (148
o
F): add 0.15 % of reading for J, K, E, and N;
and 0.45 % of reading for T]
R- and S-type: ±[0.05 % of reading + 0.4
o
C (+0.7
o
F)]
Temperature
Coefficient
0.01 % of reading + 0.03
o
C per
o
C
(0.05
o
F per
o
F) outside the specified +18
o
C to +28
o
C (+64
o
F to +82
o
F) range
[below 100
o
C (148
o
F): add 0.04 % of reading for J-, K-, E-, and N-type;
and 0.08 % of reading for T-type]
Electromagnetic
Compatibility
Susceptibility: ± 2
o
C (± 3.6
o
F) for 80 MHz to 200 MHz in 1.5 v/m field, for 200 MHz to
1000 MHz in 3 V/m field.
Emissions: Commercial limits per EN50081-1
Maximum
Differential
Common Mode
Voltage 1 V (Maximum voltage difference between T1 and T2)
Temperature
Scale ITS-90
Applicable
Standards NIST-175
Accuracy is specified for ambient temperatures between +18
o
C (+64
o
F) and +28
o
C (+82
o
F) for a period of
1 year. The above specifications do not include thermocouple error.

Other manuals for Fluke 51

Related product manuals