712 English Instruction Sheet
Page
15
General Specifications
Resolution: RTD: 0.1°C, 0.1°F. Ohms: 0.1
Maximum voltage applied between any terminal and earth
ground or between any two terminals: 30 V
Storage temperature: -20°C to 60°C
Operating temperature: -10°C to 55°C
Operating altitude: 3000 meters maximum
Temperature coefficient: 0.005% of ohms range per °C for
temperature ranges -10°C to 18°C and 28°C to 55°C. Ohms
ranges are 400 , 1.5 k, and 3.2 k.
Relative humidity: 95% up to 30°C, 75% up to 40°C, 45% up to
50°C, and 35% up to 55°C
Vibration: Random 2 g, 5 Hz to 500 Hz
Shock: 1 meter drop test
Safety: Certified as compliant to CAN/CSA C22.2 No.
1010.1:1992. Complies with ANSI/ISA S82.01-1994.
Power requirements: Single 9 V battery (ANSI/NEDA 1604A or
IEC 6LR61)
Size: 32 mm H x 87 mm W x 187 mm L (1.25 in H x 3.41 in W x
7.35 in L);
With holster and Flex-Stand: 52 mm H x 98 mm W x 201 mm L
(2.06 in H x 3.86 in W x 7.93 in L)
Weight: 337 g (11.9 oz);
With holster and Flex-Stand: 587 g (20.7 oz)