718 30G/100G
Users Manual
24
Specifications
Specifications are based on a one year calibration cycle
and apply for ambient temperature from +18 °C to +28 °C
unless stated otherwise. “Counts” are the number of
increments or decrements of the least significant digit.
Pressure Sensor Input, 718 30G
Range Accuracy
-12 to 30 psi
(-82.7 to 206.85 kPa)
±0.05 % of range
Maximum nondestructive pressure: 3X top of range (90
psi, 620 kPa, 6.2 bar)
Temperature coefficient: 0.01 % of range per
°
C for
temperature ranges -10
°
C to 18
°
C and 28
°
C to
55
°
C
Pressure Sensor Input, 718 100G
Range Accuracy
-12 to 100 psi
(-82.7 to 689.5 kPa)
±0.05 % of range
Maximum nondestructive pressure: 2X top of range
(200 psi, 1380 kPa, 13.8 bar)
Temperature coefficient: 0.01 % of range per
°
C for
temperature ranges -10
°
C to 18
°
C and 28
°
C to
55
°
C
Pressure Module Input, 718 30G and 718 100G
Range Resolution Accuracy
(determined by Pressure Module)
DC mA Input, 718 30G and 718 100G
Range Resolution Accuracy, ±(% of
Reading + Counts)
24 mA 0.001 mA 0.025 + 1
Overload protection: 125 mA, 250 V fast acting fuse
Temperature coefficient: 0.005 % of range per
°
C for
temperature ranges -10
°
C to 18
°
C and 28
°
C to
55
°
C
Loop Supply 718 30G and 718 100G
24 V dc nominal