Table 33. Required Calibration Equipment (cont.)
Calibrator
Model
Equipment Minimum Specifications
Recommended
Model
714
DC Calibrator -10 to 75 mV
Accuracy: 0.006 % + 0.25 mV
Fluke 5520A
Reference
Multimeter
-10 to 75 mV
Accuracy: 0.006 % + 0.25 mV
Fluke 8508A
TC to Banana Test
Leads
Copper Mini-Jack to Copper Wire
Type J TC Test
Lead
5500A/LEADS
715
DC Calibrator 0 to 10 Volts
Accuracy: 0.005 % + 0.5 mV
Fluke 5520A
Reference
Multimeter
0 to 10 Volts
Accuracy: 0.005 % + 0.5 mV
Fluke 8508A
Two Test Leads
717
DC Calibrator 0 to 24 mA Fluke 5520A
Reference
Multimeter
0 to 24 mA
Accuracy: 0.006 % + 0.25 mA
Fluke 8508A
Dead Weight Tester -12 to 10000 PSIG
Accuracy: 0.006 % of range
Two Test Leads 5500A/LEADS
700P Pressure
Module
Load Resistors 1 kΩ & 2 kΩ 10% 0.5W
718/718Ex
DC Calibrator 0 to 24 mA Fluke 5520A
Reference
Multimeter
0 to 24 mA
Accuracy: 0.006 % + 0.25 mA
Fluke 8508A
Dead Weight Tester -12 to 300 PSIG
Accuracy: 0.006 % of range
Two Test Leads 5500A/LEADS
700P Pressure
Module
Load Resistors 1 kΩ & 2 kΩ 10% 0.5 W
719
DC Calibrator 0 to 24 mA Fluke 5520A
Reference
Multimeter
0 to 24 mA
Accuracy: 0.006 % + 0.25 mA
Fluke 8508A
Dead Weight Tester -12 to 100 PSIG
Accuracy: 0.006 % of range
DHI PPC3
Two Test Leads 5500A/LEADS
700P Pressure
Module
Load Resistors 1.2 kΩ & 10% 0.5 W
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com

Table of Contents

Other manuals for Fluke 71X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 71X Series and is the answer not in the manual?

Fluke 71X Series Specifications

General IconGeneral
BrandFluke
Model71X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals