CONTINUITY
(-M- -Ml)
1.
Select the
RESISTANCE
function
and
desired
range
(detection
threshold
is
<10% of range
for fixed
ranges,
<20
ko
for MQ
range,
<200 for
autoranging
ko).
2.
If
visual
contin
uity in
dication is
desired,
press
[•»*
-»1
:
If
audible
continuity
indication
is
desired,
press
[•»«•
»>]
again:
(Press
again to
disable
both):
3.
Connect
the test leads as
shown
for
the
RESISTANCE
function.
4. Ensure
that
the
device being
measured
contains no
electrical
energy,
including
charged
capacitors.
WARNING
TO
AVOID
ELECTRICAL
SHOCK AND/OR
INSTRUMENT
DAMAGE,
OBSERVE
THE
FOLLOWING
MAXIMUM
LIMITS WHEN
MEASURING
CONTINUITY:
OVERLOAD
PROTECTION
500V
5.
Connect
the
test
probes
to
the
circuit
being
measured.
Continuity
between
the
test leads
will cause
the
bar
to
appear on
the display
and
the
audible
tone
to
sound
(if
enabled).
ml
FUSE
CHECK
(The A
input
contains
two
fuses.)
1.
Select
the
RESISTANCE
function
and
the 2
kfl
range.
2.
Touch the
red
test probe
to
the A
input
jack so
that the
VOS
input and
the A
input are
shorted
together.
If the
display
reads
.1000
±.0100
kn,
both fuses
are good.
If
the
display
reads
OL,
one
or both
need
replacement.
BATTERY/FUSE
TYPES
MAIN
FUSE:
2A/250V
Type
AGX2
(Inside
battery
compartment).
BACKUP
FUSE:
3A/B00V
Type
BBS-3
(Inside
Instrument
case).
Caution
Replacement
of
backup
fuse by
qualified
personnel only.
See
instruction
manual.
BATTERY:
9V
carbon-zinc
or
alkaline
NEDA Type
#1604.
BT
appears
on
display
when
approximately
20%
of
battery
life
remains.