MAINTENANCE
Performance Test
Verify the probe accuracy by measuring the output
of
a 20A
(±0.2%), 60 Hz current source (Fluke
51
OOB
and 5220A or
equivalent). When used with a compatible DMM with ±
0.75% or betterac current accuracy (Fluke 8060A or equiva-
lent), the probe should measure 20.00A ±
OAOA
(20.00 mA,
±
0040
mA).
No calibration adjustments are required.
Cleaning
Use a softclothdampened in a mild solution
of
detergentand
water to clean the 80i-600. Do not use solvents unless
absolutely necessary. A light coating
of
dripless oil on the
jaw
surfaces will prevent corrosion.