Fluke 83 - Page 16

Fluke 83
36 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
80 Series V
Calibration Manual
8
Table 11. Electrical Characteristics of the Terminals
Function
Overload
Protection
1
Input
Impedance
(nominal)
Common Mode
Rejection Ratio
(1 k unbalance)
Normal Mode Rejection
L 1000 V rms 10 M < 100 pF > 120 dB at dc,
50 Hz or 60 Hz
> 60 dB at 50 Hz or 60 Hz
F
mV
1000 V rms 10 M < 100 pF > 120 dB at dc,
50 Hz or 60 Hz
> 60 dB at 50 Hz or 60 Hz
K 1000 V rms 10 M < 100 pF
(ac-coupled)
> 60 dB, dc to 60 Hz
Full Scale Voltage Typical Short Circuit Current
Open Circuit
Test Voltage
To 6.0 M
50 M or
60 nS
600 6 k 60 k 600 k 6 M 50 M
e 1000 V rms < 7.9 V dc < 4.1 V dc < 4.5 V dc 1 mA 100 µA 10 µA 1µA 1 µA 0.5 µA
G 1000 V rms < 3.9 V dc 3.000 V dc 0.6 mA typical
10
6
V Hz max
Table 12. MIN MAX Recording Specifications
Model Nominal Response Accuracy
83 100 ms to 80 % Specified accuracy ± 12 counts for changes > 200 ms in duration
(± 40 counts in ac with beeper on)
87, 88 100 ms to 80 %
(dc functions)
120 ms to 80 %
(ac functions)
250 µs (peak)
1
Specified accuracy ± 12 counts for changes > 200 ms in duration
Specified accuracy ± 40 counts for changes > 350 ms and inputs > 25 % of
range
Specified accuracy ± 100 counts for changes > 250 µs in duration
(add ± 100 counts for readings over 6000 counts)
(add ± 100 counts for readings in Low Pass mode)
1. For repetitive peaks: 1 ms for single events.