Digital Multimeter 
   
  25 
Meter Buttons Used in the Calibration Steps 
The Meter buttons behave as follows when performing the Calibration Adjustment 
Procedure. This may be of help determining why a calibration step is not accepted and for 
determining the input value without referring to Table 16. 
A 
Press and hold to show the measured value. The measurement value is not 
calibrated so it may not match the input value. This is normal. 
B  Press and hold to display the required input amplitude. 
G  Press and hold to display the frequency of the required input. 
D 
Press to store the calibration value and advance to the next step. This button 
is also used to exit calibration mode after the calibration adjustment 
sequence is complete. 
Calibration Adjustment Procedure 
Use the following steps to adjust the Meter's calibration. If the Meter is turned off before 
completion of the adjustment procedure, the calibration constants are not changed. 
1.  While holding down B, turn the rotary knob from OFF to VAC. The Meter 
displays āZ CALā. 
2.  Press D once to see the calibration counter. 
3.  Press D again to start the password entry. The Meter displays "????". 
4.  Press 4 buttons to enter the password. 
5.  Press D to go to the first calibration step. The Meter displays "C-01" if the 
password is correct. If the password is not correct, the Meter emits a double beep, 
displays "????" and the password must be entered again. Repeat step 4. 
6.  Using Table 16, apply the input value listed for each calibration adjustment step. For 
each step, position the rotary switch and apply the input to the terminals as indicated 
in the table. 
7.  After each input value is applied, press D to accept the value and proceed to the 
next step (C-02 and so forth). 
Notes 
After pressing D, wait until the step number advances before changing 
the calibrator source or turning the Meter rotary knob. 
If the Meter rotary knob is not in the correct position, or if the measured 
value is not within the anticipated range of the input value, the Meter emits 
a double beep and will not continue to the next step. 
Some adjustment steps take longer to execute than others (10 to 15 
seconds). For these steps, the Meter will beep when the step is complete. 
Not all steps have this feature. 
8.  After the final step, the display shows "End" to indicate that the calibration 
adjustment is complete. Press D to go to meter mode.