EasyManua.ls Logo

Fluke 8508A - Temperature

Fluke 8508A
120 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Introduction and Specifications
Specifications 1
1-17
Temperature
Temperature Readout
[1] [2] [3]
Typical Equivalent Temperature Measurement Uncertainty
[12]
Resistance Range
Absolute Resistance
Measurement Uncertainty
365 day Tcal ±1 °C
[4]
±(ppm Reading + mΩ)
[11]
Probe Type
Nominal
Temperature
(°C)
Resistance
(Ω)
Accuracy
± (°C)
95% Confidence Level
0 - 199.990 000 Ω
7.5 + 0.14
25 Ω PRT/SPRT
-200 5 0.0085
25 Ω PRT/SPRT
0 25 0.0035
25 Ω PRT/SPRT
660 84 0.0025
100 Ω PRT/SPRT
-200 20 0.0035
100 Ω PRT/SPRT
0 100 0.0025
100 Ω PRT/SPRT
232 185 0.0020
200 - 1999.900 00 Ω
7.5 + 0.5
100 Ω PRT/SPRT
400 250 0.0025
99% Confidence Level
0 - 199.990 000 Ω
9.5 + 0.18
25 Ω PRT/SPRT
-200 5 0.0010
25 Ω PRT/SPRT
0 25 0.0040
25 Ω PRT/SPRT
660 84 0.0025
100 Ω PRT/SPRT
-200 20 0.0040
100 Ω PRT/SPRT
0 100 0.0025
100 Ω PRT/SPRT
232 185 0.0020
200 - 1999.900 00 Ω
9.5 + 0.6
100 Ω PRT/SPRT
400 250 0.0025
Temperature Readout (Secondary Specifications)
[1] [2] [3]
Resistance Measurement Uncertainty
Temp Coeff. ± ppm Reading/°C
Resistance Range
Transfer Uncertainty
20 Minute ±1 °C
±(ppm Reading + mΩ)
[11]
2-Wire
Adder
(Ω)
3-Wire
Adder
(Ω)
15 °C - 30 °C
5 °C - 15 °C
30 °C - 40 °C
0 - 199.990 000 Ω
0.8 + 0.14 0.1 0.005 0.6 1.0
200 - 1999.900 00 Ω
0.2 + 0.5 0.1 0.005 0.5 0.8
Type ....................................................................... 4-wire current reversal resistance measurement with readout of
equivalent temperature. 2-wire and 3-wire selectable without current
reversal. Refer to Resistance specifications for additional details.
Temperature Range .............................................. -200 °C to 660 °C, readout also available in °F or K.
Linearization ......................................................... ITS-90 or Callendar van Dusen. Entry and storage of coefficients and
nominal resistance for up to 100 probes.
Current Source ..................................................... 1 mA