4-10
Removing the A5A1 Display PCA
Use the following procedure to remove the A5A1 Display PCA. See Figure 4-7.
1. Remove the A5 Bezel Assembly (described earlier in this chapter).
2. Remove the seven small self-threading Philips screws holding the A5A1 Display
PCA to the plastic bezel (front panel). A
3. Lift the top of the A5A1 Display PCA until it clears the lip of the bezel, and then
pull the A5A1 Display PCA toward the lip of the plastic bezel. B
Before reassembling the A5 Bezel Assembly, check the position of the rubber key
pads to ensure that all of the buttons are correctly seated in their holes. C
3
1
2
avw015f.eps
Figure 4-7. Removing the A5A1 Display PCA