8808A Getting Started  Manual Supplement 
 
2/10 1   
Change #1, 44931, 46479, 47892 
On page 1, in the Specifications section, under Relative Humidity (non-condensing), replace 
Operating with the following: 
Operating ................................................ <90 % (0 C to 28 C) 
<75 % (28 C to 40 C) 
<45 % (40 C to 50 C) 
On page 5, under DC Current, Shunt Resistance, replace the Burden Voltage with the 
following: 
Burden voltage <5 mV for 200 A and 2 mA range. 
In the Input Characteristics table, change the Burden Voltage for both 200 µA and 2 mA 
ranges, 
 From: 
<1 mV 
 To: 
<5 mV 
Change #2 
On page 6, under the Warning, replace steps 4 and 5, with the following: 
4.  Insert the other end of the test lead into the mA terminal. If the fuse is good, the Meter 
displays a reading of <10 . If the fuse is blown, the Meter displays  to indicate an overload.  
5.  Remove the test lead from the mA terminal and insert it into the 10 A terminal. If the fuse is 
good, the Meter displays a reading of <2 . If the fuse is blown, the Meter displays  to 
indicate an overload. 
Change #3, 53744 
Under the LIMITED WARRANTY AND LIMITATION OF LIABILITY replace the second 
sentence with the following: 
 
The warranty period is three years limited and begins on the date of shipment. 
  On page 11, replace Table 6 with the following: 
Table 6. Accessories 
Item  Model / Part Number 
 Premium Test Lead Set  TL71 
Fuse, .25*1.25, 0.063 A, 250 V, Slow  163030 
Fuse, .25*1.25, 0.125 A, 250 V, Slow  166488 
F1 - Fuse, 11 A, 1000 V, Fast, 406INX1.5IN, BULK  803293 
F2 - Fuse, 440 mA, 1000 V, Fast, 406INX1.375IN, BULK  943121 
Rack Mount Kit 8845A & 8846A Single   Y8846S 
Rack Mount Kit 8845A & 8846A Dual  Y8846D 
USB to RS-232 Cable Adapter (included)  884x-USB 
Precision Electronic Probe Set  TL910 
2X4 Wire Ohms 1000 V Test Lead  TL2X4W-PTII 
FlukeView Forms Basic Software (included)  FVF-SC5 
FlukeView Forms Software Upgrade to enhanced version  FVF-UG