8845A/8846A
Calibration Manual
A-4
Table A-1. Blank Verification Record for 90-Day Specifications (cont)
Nominal Output Test Limits
Ampl. Freq.
Range
High Low
Results Pass/Fail
8846A Only AC Volts (cont)
320 20000 1000 320.460 319.540
320 50000 1000 320.852E+0 319.148E+0
320 100000 1000 322.72E+0 317.28E+0
8845A Only AC Volts
750 45 750 750.600 749.400
750 1000 750 750.600 749.400
750 10000 750 750.600 749.400
320 20000 750 320.385 319.615
320 50000 750 320.727E+0 319.273E+0
320 100000 750 322.52E+0 317.48E+0
8845A/8846A Frequency
1 10 10.00300 9.99700
1 40 40.01200 39.98800
0.1 300000 300030.0 299970.0
8846A Only Frequency
0.1 1000000 1000100.0 999900.0
8846A Only 4-Wire Resistance
0 10 3.0E-3 0
10 10 10.00380 9.99620
8845A/8846A 4-Wire Resistance
0 100 4.0E-3 0
100 100 100.0120 99.9880
0 1000 10.0E-3 0
1000 1000 1000.090 999.910
0 10000 100.0E-3 0
10000 10000 10000.90 9999.10
0 100000 1.000000 0
100000 100000 100009.0 99991.0