11
Specications and Environmental Conditions
Environmental Conditions
temperature range: 5-40°C (41-104°F)
ambient relative humidity: maximum 80% for temperature <31°C, decreasing
linearly to 50% at 40°C
pressure: 75kPa-106kPa
mains voltage: within ±10% of nominal
vibrations in the calibration environment should be minimized
altitude: less than 2,000 meters