Wireless K-Type Thermometer
Required Equipment
5
Required Equipment
The equipment in Table 2 is necessary for performance tests and calibration adjustment.
Table 2. Required Equipment
Equipment Required Characteristics
Recommended
Model
Notes
Calibrator
Output Voltage: 30 mV dc
Accuracy: ±0.125 %
Resolution: 0.1 µA
Fluke 5500A Calibrator
(or equivalent)
Calibration only
Temperature Accuracy
Verification only
-25 to 100 °C
-200 to 1300 °C
±0.16 °C
±0.40 °C
Copper Wire Calibration only
Thermocouple
Connector
Copper Fluke PN 601747 Calibration only
Thermocouple
Connector
For K-type thermocouple
Fluke 80CK-M
PN 779942
Verification only
K-type
Thermocouple
Wire
Standard Omega Verification only
Use copper wire, not thermocouple wire, to source voltage during calibration. Use copper plugs with
copper wire. Use yellow K-type plugs and wires to source temperature.