Before Calibration Adjustment
Before the Product calibration can be adjusted, you must go through the maintenance
mode menu and enter your password.
Maintenance Mode
The Product maintenance mode can be used to set different parameters on the Product
that include auto power off, backlight adjustment, and calibration. To use the
maintenance mode:
1. With the Product off, push and hold .
2. Push . Keep pushed until all the display segments are shown.
3. Release and .
The Product is now in maintenance mode.
Password Entry
To go to the calibration mode, push five times until CAL is shown. You will need to
enter a password to access calibration mode.
To enter the password:
1. Push and the CAL version is shown. For example n002.
2. Push to show “????”.
3. Push to change the first digit of your password (default: 1234).
4. Push to confirm your choice. The subsequent “?” changes to “_”.
5. Do steps 3 and 4 again to enter the subsequent digits of the four-number password.
6. When all of the correct digits are entered, push to confirm the input.
If the correct password is entered, “C-01” is shown. If the incorrect password is
entered, “????” is shown and the password must be correctly entered.
Change the Password
Note
If you change the password and then lose it, see the “Restore the Default
Password” section.
To change the password:
1. Do steps 1 through 5 in the “Password Entry” section.
2. Before you push to confirm your final input (step 6), push to show “----” on
the display.
3. Push to change the first “-” to the first digit of your new password.
4.
Push to co
nfirm your ch
oice.
5. Repeat steps 3 and 4 to enter the subsequent digits of the new four-number password.
6. When the correct digits are entered, push to confirm the input and change the
password. If the Product has been calibrated, it will go to normal measurement
mode.