SmartPaq calculations will be performed on data from all the
thermocouples that are specified in the SmartPaq setup (see below).
The SmartPaq result will be a fail if:
One or more of the chosen parameters (see below) is outside the
limits specified for any of the thermocouples .
Data from any of the specified thermocouples contains invalid data (e.g.
open circuit, see p. 67).
SmartPaq Setup
Parameters for the SmartPaq calculations are programmed into the logger as
part of the reset procedure (p. 46). In Insight’s Logger Reset dialog, click the
SmartPaq button to open the SmartPaq Setup dialog and specify or edit your
parameters as follows.
For identification purposes, you may enter a description for your new set of
parameters. When you have defined (or edited) them, click Save As to save
these (as a SmartPaq parameters file, *.PFP) for future use; or click Open to
use a previously-saved set. The description is included in the report of the
profile run which you can print after downloading the data (see p. 61).
You can specify parameters and limits for any combination of:
Datapaq Value (one cure schedule).
Time at Temperature (up to three thresholds with Oven Tracker
Professional, one threshold with Over Tracker Basic).
Maximum Temperature.
For details of these analysis modes as used within Insight, select ‘Data
Analysis’ in the online Help system (on Insight’s menu bar, select Help > Contents).
Select the analyses which are to form part of the SmartPaq calculation.
NB Failure to pass any one of the selected analyses will cause an
overall fail in the SmartPaq calculation (see above).
Datapaq Value
Enter cure-schedule data to suit your process, as recommended by your
coating manufacturer:
Temperature and time data for three cure options.
Minimum and maximum permitted temperatures for successful cure.
OVEN TRACKER XL2 Running a Temperature Profile 51