©Fluke Calibration 2011 Basics Of Measurement Uncertainty for DMM Calibration 30
Type A uncertainty is determined by the statistical
analysis of a series of observations (measurements).
Type A uncertainties includes effects from:
Variations of multiple repeated readings from the UUT
Effects of the system noise
Noise and short term variation of the standard
Now let’s examine the basic statistics …