©Fluke Calibration 2011 Basics Of Measurement Uncertainty for DMM Calibration 65
“Automation”
some words of wisdom
Remember, it is always the metrologist’s responsibility to
insure proper calculation of measurement uncertainty
Every lab has unique characteristics which must be supported
Configuring the measurement characteristics is also unique
Defining the specific error budget for the test
Configuring the specific measurement uncertainty parameters
There should be definite information to support answering
any auditor’s questions
Keep records of the procedure’s measurement design with
an uncertainty error budget
Be able to demonstrate the reasonableness of the test’s
uncertainties