EasyManua.ls Logo

Fluke SUPER-DAQ 1586A - Thermistor Verification Steps

Fluke SUPER-DAQ 1586A
38 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1586A
Calibration Manual
12
Thermistor Verification Steps
Connect the Product to the test equipment as shown in Figure 5 and apply the
resistance listed in Table 6. Configure Ch001 as a 4-wire thermistor function with
resistance display.
5522A CalibratorUUT
NC
NC
hcn204.wmf
Figure 5. Thermistor Test Equipment Setup
Table 6. Thermistor Verification Steps
Nominal
Input
Range
1-Year Test Limits
High Low
0 Ω 2.2 k 0.2 Ω -0.2 Ω
2 kΩ 2.2 k 2000.28 Ω 1999.72 Ω
90 kΩ 98 k 90004.1 Ω 89995.9 Ω
900 kΩ 1 M 900140 Ω 899860 Ω

Related product manuals