SUPER-DAQ Precision Temperature Scanner
Calibration Adjustment
23
Table 9. Adjustment Steps (cont.)
Step Value Range
Input
Signal
Description
Entry
Point
DCI
DC100UA 100.0E-6
Start calibration of dc
100 μA range
Y
DC_P100UA_1 100.0 μA N
DC_0A_1 Open N
DC_N100UA_1 -100.0 μA N
DC1MA 1.0E-3
Start calibration of dc 1 mA
range
Y
DC_P1_0MA_2 1.0 mA N
DC_0A_2 Open N
DC_N1_0MA_2 -1.0 mA N
DC10MA 10.0E-3
Start calibration of dc
10 mA range
Y
DC_P10MA_3 10.0 mA N
DC_0A_3 Open N
DC_N10MA_3 -10.0 mA N
DC100MA 100.0E-3
Start calibration of dc
100 mA range
Y
DC_P100MA_4 100.0 mA N
DC_0A_4 Open N
DC_N100MA_4
-100.0
mA
N