1586A
Calibration Manual
24
Remote Commands for Calibration
Table 9 alphabetically lists the command set for calibration.
Table 9. List of Commands
Remote Command Meaning
CALibrate:ABORt
Instruct Product to abort calibration procedure after
present step.
CALibrate:BACKup
Backup to previous entry point in calibration
procedure.
CALibrate:CONSt?
Retrieve the value in use of the given calibration
constant.
CALibrate:DATE?
Return a CAL date associated with stored
calibration constants.
CALibrate:INFO?
Return message or instructions associated with
running step.
CALibrate:MOD:DATE
<slot>,<year>,<month>,<day>
Set CAL date of module in specified slot.
CALibrate:NEXT [<reference>] Continue a calibration procedure if it is waiting.
CALibrate:REF? Return nominal value expected for reference entry.
CALibrate:SECure:STATe
<boolean>,<admin_password>
Instruct unit to enable calibration.
CALibrate:SECTion Skip to next section of calibration procedure.
CALibrate:SKIP Skip to next entry point in calibration procedure.
CALibrate:STARt <procedure>[,<step>] Start a calibration procedure.
CALibrate:STATe? Return state of calibration.
CALibrate:STEP? Return name of step currently running.
CALibrate:STORe Store new calibration constants.
CALibrate:MODule:DATE
<slot>,<year>,<month>,<day>
Set module CAL date.