11. Troubleshooting
Malfunction Cause Solution
LCD display: No display
Loose battery contacts Check batteries for good
contact
Accuracy of measurements
is insucient
K FACTOR wrong
Meter operates below
admissible min. ow rate
Check K FACTOR,
a
see chapter 8.3
Increase ow rate to
acceptable level
Reduced ow rate,
or no ow at all
Oval gear blocked
Clean measuring chamber
Display ashing „Err 1“
Data stored in electronic
memory are damaged
Sorry, irreparable
Short display of „Err 2“
Limited error reading data
(usually when exchanging
batteries)
Control circuit board automa-
tically switches on and o to
restore proper operation
12. Repairs/service