EasyManua.ls Logo

Focus 4010 - Page 13

Focus 4010
26 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
FOCUS 4010/HT Dental Porcelain Furnace Page 13 Manual version 4.2.2013
In case one or couple of the parameter fields are not in use, they should remain
blank and there is no need to introduce the “0level or any value in the blank field.
EXAMPLE 1
Entering a START TEMP of 400C and HEAT RATE 55C/min .
1. Press the Arrow key until the [START TEMP] field will appear
2. Press the keys [4], [0], [0].
3. Press the Arrow UP key until the HEAT RATE display will be illuminated (do not
enter values in the UP TIME and DRY TIME fields if not needed).
4. Press the keys [5], [5] for the ramp.
5. Press ENTER key.
EXAMPLE 2
Entering a UP TIME and DRY TIME of 3.5 minutes each.
1. Press the arrow key until the [UP TIME] field will appear
2. Press the keys [3], [3], [0].
3. Press the up key.
4. Press the keys [3}, [3], [0].
5. Press the key [ENTER] or Arrow key again.
EXAMPLE 3
Entering a “VACUUM ON TEMP of 420C, “VACUUM OFF” TEMP 820C and
VACUUM LEVEL of 720mmHg.
1. Press the up key until the Field "V on" will be displayed on the “VACUUM field.
2. Press the keys [4], [2], [0]. For vacuum start
3. Press the up key until the field " vacuum off" will be displayed on the “V.off
display. For the end of the vacuum
4. Press the keys [8], [2], [0].
5. The vacuum Level can be set here or by preset the Level in the option screen, if
the Level is chosen this possibility will not appear, if the fields in the option is
empty, you will be able to set the vacuum level
if the field is enable than
6. Press the keys [7], [2], [0].
7. Press the key [ENTER] or arrow key again.
1.