SPECIFICATIONS
System Performance
Power Source: Four 1.5V AA alkaline batteries
Size of Monitor w/o Cuff:
160 mm (L) x 95 mm (W) x 66.5 mm (H), around 400g without battery
Cuff Size:
Wide range 9.4-16.9 inches (24-43 cm)
Memory: Maximum 60 memory records
Power Saving: Automatic power off if system is idle for 3 minutes
System Operating Condition:
41°F to 104°F (5°C to 40°C), 15 - 93% RH
Meter Storage Condition:
-13°F to 158°F (-25°C to 70°C), 10 - 95% RH
Power Supply Input: DC +6V / 1A (max) via Power Plug
Expected Service Life: 3 years
Blood Pressure Measurement Performance
Measurement Unit: mmHg
Systolic Measurement Range: 60 to 255 mmHg
Diastolic Measurement Range: 30 to 195 mmHg
Pulse Rate Measurement Range: 40 to 199 beats / minute
Accuracy of Pressure: ±3 mmHg or ±2% of reading
Accuracy of Pulse Rate: ±4% of reading
This device has been tested to meet the electrical and safety requirements of: IEC/EN
60601-1, IEC/EN 60601-2.
Reference to Standards:
• IEC 60601-1 General requirement for safety
• IEC 60601-1-2 Requirements for EMC
• AAMI/ANSI /IEC 80601-2-30, ANSI/AAMI/ISO 81060-2, NIBP requirements