Emphasizes avoiding maintenance with patients connected and using OEM parts for safety and warranty.
Warns about servicing by qualified personnel and proper electrical connections to prevent shock.
Tests automatic alarms and pressure holding capabilities to ensure proper machine response.
Tests the machine's audible alarm function when the main power supply is turned off.
Detailed procedures for rebuilding the ultrafiltration, concentrate, and bicarbonate diaphragm pumps.
Inspects the heater element for corrosion and checks its resistance for proper operation.
Re-checks high voltage AC connections, power cord, strain relief, and wiring for safety.
Conducts the Online Clearance Test (OLC) and troubleshooting steps if the machine is equipped.
Tests the volume pumped by concentrate and bicarbonate pumps, comparing to expected values.
Checks the 5-volt supply status on the debug screen for correct operation and tolerance.
Verifies blood leak and dimness sensor readings in volts are within the specified operational limits.
Calibrates arterial, venous, and transmembrane pressure readings, checking for leakage.