What to do if Fresenius Medical Care 2008K Medical Equipment shows Venous Pressure Alarm (with the upper Venous Pressure Alarm limit flashing)?
- KKatie MartinezAug 25, 2025
If high pressure is detected in the venous drip chamber: 1. Check venous tubing for loose connections, kinks, clotting, or loose clamps. 2. Ensure that the transducer protector is dry and the monitor line is open. Replace the transducer protector, if necessary. 3. Check access point for clotting and needle position. 4. Press RESET to reset alarm. Press the RESET key again and hold for one second to select new alarm limits. If the condition persists, reduce the blood flow rate. If the alarm won’t reset, take the machine out of service and alert a qualified technician.