What should I do if the TMP is Low (alarm at or below 60) on my Fresenius Medical Care Medical Equipment?
- TTimothy NelsonSep 13, 2025
If the TMP is low (at or below 60), here’s what you should do: 1. Make sure the venous transducer protector is dry and the monitor line is open. Replace the protector if needed. 2. Inspect the dialyzer supply and return lines for any kinks. 3. Check the filter screen in the dialyzer return line to ensure it's clean. 4. Press the RESET key to reset the alarm. Press the Override key and hold for one second to select new alarm limits or adjust the TMP. If you still can't reset the alarm, return the blood to the patient if the alarm occurs during treatment. Then, take the machine out of service and contact a qualified service technician.