What to do if Fresenius Medical Care 2008K displays Venous Pressure Alarm (with the lower Venous Pressure Alarm limit flashing)?
- OOlivia WhiteAug 27, 2025
If low pressure is detected in the venous drip chamber: 1. Check venous tubing for kinked line, clotting or clamps. 2. Check for a disconnected line. 3. Ensure that the transducer protector is dry and the monitor line is open. Replace the transducer protector, if necessary. 4. Press RESET to reset alarm. Press the RESET key again and hold for one second to select new alarm limits. If you are unable to reset the alarm, take the machine out of service and alert a qualified technician.