PatientCard Reader
Installation Instructions
PatientCard Reader Installation Instructions
for
2008T BlueStar
Hemodialysis Machines
Part Number 490330 Rev B
This document contains proprietary information of Fresenius USA, Inc. d/b/a Fresenius Medical Care North America
and its affiliates (“Fresenius Medical Care”). The contents of this document may not be disclosed to third parties,
copied, or duplicated in any form, in whole or in part, without the prior written permission of Fresenius Medical Care.
Fresenius Medical Care, the triangle logo, 2008, BlueStar, bibag and DIASAFE are trademarks of Fresenius Medical
Care Holdings, Inc., or its affiliated companies. All other trademarks are the property of their respective owners.
If there are any questions regarding these instructions, contact Fresenius Medical Care Technical Support at 800-227-2572.
© 2019, Fresenius Medical Care North America. All rights reserved.
Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T BlueStar and is the answer not in the manual?

Fresenius Medical Care 2008T BlueStar Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T BlueStar
CategoryMedical Equipment
LanguageEnglish

Summary

Software and System Preparation

Software Verification

Verify machine software versions meet installation requirements for compatibility.

Machine Disassembly and Component Preparation

IV Pole Removal

Remove the existing IV pole holder to make way for the new component.

Card Cage Preparation

Prepare the card cage by removing screws for access to internal components.

Installation of Card Reader Components

Card Reader Adapter Board Assembly Installation

Install the adapter board and connect USB and serial cables to the assembly.

Card Reader/Upper IV Pole Holder Installation

Install the new IV Pole Holder assembly with the card reader and connect the USB cable.

Machine Reassembly and Configuration

Machine Reassembly

Reassemble the machine by reinstalling covers, IV pole, and securing components.

Card Reader Option Configuration

Enable the Patient Card option in the machine's Service Mode settings.

Post Installation Testing

PatientCard Functionality Test

Perform tests to verify the PatientCard reader functions correctly after installation.