EasyManuals Logo

Future Design LIMIT L91 User Manual

Default Icon
52 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #46 background imageLoading...
Page #46 background image
Step 4: Calibrate of .
Setup the equipment same as step 3.
gain cold junctiongain cold junction
46
Step 5: Calibrate .
Send a 100 ohms signal to terminal 3, 4 and 5 according to
RTD reference voltage
Figure 5-3
.
RTD reference voltage
Figure
5-3
1
2
3
4
5
6
7
8
9
10
100 ohms
Figure 5-3
RTD Calibration
Figure 5-3
RTD
Calibration
Press for at least 4 seconds. The display will blink a
moment. If the display didn't blink, then the calibration failed.
Step 6: Calibrate .RTD serial resistanceRTD serial resistance
Change the ohm's value of the calibrator to 300 ohms. Press
for at least 4 seconds. The display will blink a moment.
If the display didn't blink, then the calibration failed.
UM L91-Rev 8
The L91 being calibrated for Cold Junction Compensation MUST
be programmed for K t/c input, Celsius display to performing
the CJG calibration.
prior
NOTE: The unit under calibration must powered at an
ambient temperature of 50C +/-3. Allow at least 20 minutes
for warming up.
NOTE: The unit under calibration must powered at an
ambient
temperature of 50C +/-3. Allow at least 20 minutes
for
warming up.
Set the calibrator to be configured as K type thermocouple
output. Calibrator must have an internal compensation.
Send a 0.00 C signal to the unit under calibration.
With CJG on the display adjust the value to 0.00 reading.
Once adjusted, Press for at least 4 seconds. The display will
blink a moment. If the display didn't blink, then the calibration
failed.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Future Design LIMIT L91 and is the answer not in the manual?

Future Design LIMIT L91 Specifications

General IconGeneral
BrandFuture Design
ModelLIMIT L91
CategoryController
LanguageEnglish

Related product manuals