- 20 -
APPENDIX II
Conducted
disturbances
induced by RF
fields
IEC 61000-4-6
3 V
rms
a
0.15 MHz - 80 MHz
6 V
rms
a
in ISM and
amateur radio
bands between
0,15 MHz and 80
MHz
b
80% AM at 1 kHz
3 V
rms
a
0.15 MHz - 80 MHz
6 V
rms
a
in ISM and
amateur radio
bands between
0,15 MHz and 80
MHz
b
80% AM at 1 kHz
Radiated RF EM
fields
IEC 61000-4-3
10 V/m
c
80 MHz - 2.7 GHz
80% AM at 1 kHz
10 V/m
c
80 MHz - 2.7 GHz
80% AM at 1 kHz
The Sphygmomanometer
(
MD2200
)
is intended for use in the electromagnetic environment specified below. The customer
of the user of Sphygmomanometer
(
MD2200
)
should assure that it is used in such an environment.
Portable and mobile RF communications equipment should be used
no closer to any part of the Sphygmomanometer (
MD2200
),
including cables, than the recommended separation distance
calculated from the equation applicable to the frequency of the
transmitter.
Recommended separation distance
where P is the maximum output power rating of the transmitter in
Watts (W), d is the minimum recommended separation distance in
meters (m), and E is the immunity test level in V/m.
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey,
a
should be less than the compliance
level in each frequency range.
Interference may occur in the vicinity of equipment marked with the
following symbol:
Immunity test Compliance level Electromagnetic environment - guidance
IEC 60601
test level