Materials Needed:
MIC TRAC MT-3000 base unit and CPU readout
5/32" hex wrench
Seiko printer (optional)
Lightweight gage oil
Depth Micrometer
Gage Calibration Record
1. Turn the fine adjust knob clockwise to bring the
depth micrometer face in contact with the flat
face anvil.
2. Verify that the readout on the CPU displays
0.00000.
3. Loosen the coarse adjust lock and turn the
coarse adjust knob clockwise to slide the
movable carriage to the right.
4. Adjust the depth micrometer to the first
calibration value.
Note: If you pass the calibration value, turn the
spindle back and approach the value again,
from the same direction. This practice will
increase the accuracy of the calibration.
the MT-3000, go to page 11.
Calibration

Other manuals for GAGEMAKER MIC TRAC MT-3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GAGEMAKER MIC TRAC MT-3000 and is the answer not in the manual?

GAGEMAKER MIC TRAC MT-3000 Specifications

General IconGeneral
BrandGAGEMAKER
ModelMIC TRAC MT-3000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals