32
The mixing process using the UNGUATOR
®
e/s
Fault Fault remedy Possible causes of fault
display
FE 0 Hit the “Start/Stop” button, Switching the device on and off during
oscillation arm moves to a work phase or temporary blackout
the next end position
FE 1 Remove jar and change-over between the operating modes,
hit the “Start/Stop” button, when a mixing unit has locked in place
Oscillation arm moves to or switching on the device when a mixing
the desired position mixing unit has locked in place
FE 2 Hit the “Start/Stop” button, Oscillation arm does not move or very
FE 0 displays slowly
FE 8 Hit the “Start/Stop” button, No mixing unit locked in place when in
snap in jar - reboot “HAND” mode
FE 9 Oscillation arm will automatically No mixing unit attached to oscillation arm
return to home position in “AUTO” mode
FE 9 Remove jar and No mixing unit snapped in or released
hit the “Start/Stop” button in “AUTO” mode
FE A Hit the “Start/Stop” button, Temporary motor overload
possibly switch off device
Table 3-5: UNGUATOR
®
e/s error messages
Error messages will also occur if UNGUATOR
®
MBs are used that do not match the
respective size of the UNGUATOR
®
Jar or if the end position of the previous mixing
process has been left at the beginning of the mixing process rather than the start
position required for the new mixing process.
Incorrect operation or environmental influences may cause problems not listed in the
error list. If this occurs, switch off the device or pull the power plug. The UNGUATOR
®
e/s
may then be switched on again after period of 10 seconds.