Venous Pressure Alarms
Appears:
The venous pressure has reached the set venous pressure
high alarm limit.
Machine actions:
- The blood pump stops.
- The arterial blood line clamp closes in single needle
mode.
- The ultraltration rate is set to zero.
When the venous pressure is within the set alarm limits
again, the blood pump automatically starts.
Possible cause:
- The venous blood line is kinked or clamped.
- The position of the venous needle has changed.
- Clotting has occurred in the venous blood line after
the venous pressure measurement point; in the blood
line, in the venous drip chamber, or in the needle.
1
3:14
AK 96
®
Operator's Manual - Alarms
HCEN12239 Revision 12.2010
Program version 3.xx