36
2) Connect the cell cable to the correct color-coded receptacles on the 20 Calibration Cell.
3) Place the Calibration Cell inside the Calibration Shield, close the lid, and connect the black
floating-ground lead of your cell cable to the Shield’s grounding post. (Note that the photograph
below left is of a 200 cell, so is only a general representation of the procedure.)
4) Open Gamry Framework™ software. Select Experiment > Named Script...
The Select a script to run window appears.
5) From the list of scripts, choose
calcable.exp, then click the Open
button.
The Cable Capacitance Calibration window
appears.