For example:
Parameter Meaning
0.1 0.1 milliseconds
200us 200 microseconds
0.1s 0.1 seconds
100ma 100mA
2.45A 2.45A
2.3 2300mA or 2.3A
Note that parameters are in “USA/UK” format so that a half is written “0.5” not “0,5”
The command codes and their meaning are described below. The upper case commands are
shown, followed by lower case letters denoting the numeric argument.
Error number Reason
Err 1 A parameter value is invalid
Err 2 Command not recognised
Err 3 Numeric value is wrong format
Err 4 Wrong number of parameters
Any changes made using Ethernet commands are not saved permanently until the AW
command has been issued.
12.3.1 General Commands
Save the settings to memory.
AW
Once the settings are saved to memory they are then retained when the unit is switched off. If
this is not done, changes to the settings are volatile, and if the unit is switched off they revert to
those in force when the last AW command was issued.