T/PUMP
®
TP600 & 700 Series Localized Temperature
Therapy System
Service Manual
P/N 101176000 Rev B 9/11

Questions and Answers

  • M
    Mark HayesAug 10, 2025
    What to do if my Gaymar T/PUMP TP700 T/Pump will NOT turn on?
    • P
      Phillip CampbellAug 10, 2025
      First, ensure the electrical cord is fully inserted into a properly grounded Hospital Grade receptacle. Then, visually inspect the power cord for any defects and replace it if necessary. If these steps don't resolve the issue, the membrane panel may be defective and need replacement. For a defective PC Board, refer to the troubleshooting steps for when the T/Pump will NOT heat.
  • A
    Adam BrownAug 17, 2025
    What to do if my Gaymar T/PUMP TP700 T/Pump will NOT pump?
    • M
      melissa38Aug 17, 2025
      First, check that the water level is adequate; refill the reservoir with room temperature water to the proper level if it's low or empty. Next, check for power to the motor (TP650(C)/ TP700(C) - 120V ± 10% or TP702 - 230V ± 10%) across the motor connections. If power is present, inspect the motor, impeller magnet, and magnet driver assembly for defects. Replace the motor if the fan isn't spinning when plugged in and in run mode. Check the impeller for obstructions or replace the impeller magnet assembly. Ensure the screw holding the driver magnet to the motor shaft is tight.
  • T
    trodriguezAug 25, 2025
    Why is the warning indicator ON when my Gaymar T/PUMP TP700 Medical Equipment is in Standby Mode?
    • Y
      Yolanda MooneyAug 25, 2025
      The unit likely shut down due to an over-temperature condition. To resolve this, empty the reservoir and refill it with room temperature water. Ensure all clamps are open. Then, press the On/Standby button and verify flow through the pad. The warning light should turn off within 5 minutes.
  • M
    Mr. Joseph HenrySep 1, 2025
    How to unlock temperature or therapy time buttons on Gaymar Medical Equipment (TP700 Series Only)?
    • X
      xhamiltonSep 1, 2025
      If the temperature or therapy time buttons on your TP700 series Gaymar Medical Equipment are not working, it's possible they have been locked. To unlock them, simply press and hold the lock button for two seconds. If this does not work, the membrane panel may be defective and needs to be replaced.

Summary

SAFETY PRECAUTIONS

DANGER

Critical safety warnings for immediate hazards.

WARNING

Important safety precautions to prevent injury or damage.

CAUTION

Advisories to prevent minor damage or operational issues.

SPECIFICATIONS

START-UP PROCEDURE

SHUTDOWN PROCEDURE

WATER TEMPERATURE CONTROL

THEORY OF OPERATION

FUNCTIONAL CHECK

DISASSEMBLY;REASSEMBLY

Reservoir

Steps for disassembling the T/Pump reservoir.

TROUBLESHOOTING

Gaymar T/PUMP TP700 Series Specifications

General IconGeneral
ModelTP700 Series
ApplicationTherapeutic heating and cooling
Over Temperature CutoutYes
Sleeve OptionsMultiple sizes available
PortabilityPortable
DisplayDigital display
Alarmstemperature