What causes NIBP air leakage in GE HEALTHCARE E-PSM?
- CCrystal OliverAug 26, 2025
If the GE HEALTHCARE Medical Equipment is showing 'NIBP air leakage', it could be due to a leaking hose or cuff, in which case you should replace the cuff. If the cuff connector is damaged, replace it, especially if the fault is in the hose connector. Check for a damaged or missing O-ring and replace it if necessary. Also, inspect the hose double connector for damage and replace the NIBP cuff hose if needed. Examine the tubes connecting the patient connector panel and the module for leakage, and check the tubes inside the module, replacing the whole tubing if there's leakage. If a tube is disconnected or damaged, fix the connections. Finally, if the manifold is leaking or the tubes/valves are damaged, replace them.