5-20 ApexPro™ 2001989-351A
Troubleshooting
NOTE
“XXX” is representative of the number of events stored in the error log. It
will appear on screen as an actual number.
7. At the @> prompt type del XXX (where XXX equals the number of events as
indicated in the error message) and press Enter to display the most recent event.
Other commands include:
Type , (comma) and press Enter to display the previous log entry.
Type . (period) and press Enter to display the next log entry.
Type dael and press Enter to dump all entries. This is useful in conjunction with
the “capture-to-file” feature of Hyperterminal.
The following is an example of the information in each event log.
The date and time are set by the ApexPro Telemetry System. If the date and time have
not been updated by the time the event occurs, the time defaults to 1 January 1990
00:00.
The error text explains the error code.
The severity levels are:
4 = An event occurred, not an error
5 = A minor error
6 = An error that caused the system to halt
7 = An error that caused the system to reset
The number of ticks indicates when the event occurred relative to the last time the
system was rebooted. There are 60 ticks in one second.
Record Number 244
Date/Time: 1 January 2000 00:00
Error Code: 0x438a
Severity: 4
Error Text: Set fields for rcvr 15 to 15
Process Name: sysstart
Status Register: 0x9042
User Stack Pointer: 0xff0902c8
Program Counter: 0xff013a24
Super Stack Pointer: 0xffcbffe8
Heap Pointer: 0xff0c8554
Passed Parameter: 15
Error Number: 60544
Ticks: 525