Do you have a question about the GE AMX 4+ and is the answer not in the manual?
Type | Mobile X-ray System |
---|---|
Maximum kVp | 125 kVp |
Collimator | Manual Collimator |
Battery | Rechargeable, lead-acid |
Image Receptor Compatibility | CR and DR systems |
Focal Spot Size | 1.2 mm |
mAs Range | 0.04 - 320 mAs |
X-ray Tube | Rotating Anode |
Identifies the AMX-4+ by model number and lists part/catalog numbers.
Overview of AMX-4+ safety safeguards and reference to operating manuals.
Guidelines for frequent cleaning of the AMX-4+ to maintain performance.
Detailed steps for cleaning and disinfecting the hand switch safely.
Requirements and recommendations for periodic inspection programs.
Performance requirements for diagnostic x-ray equipment from HHS.
List of required tools and materials for service and maintenance.
Perform defined checks as per Direction 2173224-100.
Perform defined checks on the power cord as per Direction 2173224-100.
Inspect main, safety, and follower counterweight cables per Direction 2190775-100.
Verify collimator movement and mounting screw tightness.
Adjust drive belt tension and torque specifications as per Illustration 2-2.
Perform HHS tests excluding AEC as per Direction 2173221-100.
Perform AEC HHS tests as per Direction 2173221-100.