AMX4+ PERIODIC MAINTENANCE
GE Healthcare MODELS: 2169360, 2236420 & 2275938
REV 10 DIRECTION 2173227100
2
Satisfactory equipment performance requires the use of service personnel specially trained on xray apparatus. GE
Healthcare is responsible for the effects on safety, reliability, and performance only if the following conditions are
met:
The electrical wiring of the relevant rooms complies with all national and local codes.
All assembly operations, extensions, readjustments, modifications, or repairs are carried out by GE
Healthcare authorized service representatives.
The equipment is used in accordance with the instructions for use. Refer to Direction 21669131EN,
AMX4+ Operating Manual, or Direction 2166911100, AMX4+ International Operation, for proper
operating procedures.
Only trained and qualified personnel should be permitted access to the internal
parts of this equipment.
GE Healthcare, and its associates, maintain a worldwide organization of stations from which one may obtain
skilled xray service. If desired, arrangements can usually be made to furnish preventive and/or emergency
service on a contract basis. A GE Healthcare representative will be glad to discuss this plan.
1-3 Cleaning
The AMX4+ should be cleaned frequently, particularly if corroding chemicals are present. Use a cloth slightly
moistened in warm, soapy water to clean all surfaces. Wipe with a cloth slightly moistened in clean water. Do not
use cleaners or solvents of any kind. They may remove markings and could damage the finish and plastic covers.
1-3-1 Hand switch Cleaning and Disinfecting Instructions
SAFETY PRECAUTIONS Before you Begin
· Disconnect the hand switch cord from the body of the hand switch before performing the maintenance / cleaning
procedures
· Never use solvents or flammable solutions to clean the hand switch
· Never use a dripping cloth (or) immerse hand switch in water or cleaning solutions
INSTRUCTIONS
Use a cloth moistened in warm soapy water (use mild soap) to clean the hand switch.
APPROVED CLEANERS
The cleaners listed below are approved for cleaning the hand switch:
· Bleach 50% mix with water (58% household Bleach)
· Glutaraldehyde <5%
· Polyethylene Glycol <20% (tested as Cidex Plus 28)
· Isopropyl Alcohol 70% concentration
· Hydrogen Peroxide 1540% concentration
Never use cleaners or solvents of any kind if you are uncertain of the nature of the
cleaning agent. The hand switch should be cleaned using EPA cleared and EPA
registered highlevel disinfecting agents.

Other manuals for GE AMX 4+

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE AMX 4+ and is the answer not in the manual?

GE AMX 4+ Specifications

General IconGeneral
TypeMobile X-ray System
Maximum kVp125 kVp
CollimatorManual Collimator
BatteryRechargeable, lead-acid
Image Receptor CompatibilityCR and DR systems
Focal Spot Size1.2 mm
mAs Range0.04 - 320 mAs
X-ray TubeRotating Anode