A-10 ApexPro™ 2001989-351A
Technical specifications
Pressure: Dynamic or fixed programmable. Dynamic pressure
configuration maximizes patient comfort by limiting cuff
inflation to 30 mmHg greater than the previous systolic
pressure, without exceeding configurable limits.
Pressure Range: 0 to 250 mmHg
Maximum pressure
programmable limits:
100 to 250 mmHg, increments of 10 mmHg
Minimum pressure
programmable limits:
10 to 100 mmHg, increments of 10 mmHg
Auto retry: Automatically initiates an additional reading in the event a
reading fails to satisfy any of the programmable criteria.
Blood Pressure Range *Contact technical support for information on setting these
limits.
Systolic: 10 to 250 mmHg
Upper auto-retry
programmable limits:
50 to 240 mmHg, increments of 10 mmHg*
Lower auto-retry
programmable limits:
50 to 150 mmHg, increments of 10 mmHg*
Systolic change (delta)
limits:
30 to 100 mmHg, increments of 10 mmHg*
Diastolic: 10 to 250 mmHg
Upper auto-retry
programmable limits:
50 to 150 mmHg, increments of 10 mmHg*
Lower auto-retry
programmable limits:
30 to 100 mmHg, increments of 10 mmHg*
Diastolic change (delta)
limits:
20 to 100 mmHg, increments of 10 mmHg*
Pulse Pressure
Upper auto-retry
programmable limits:
40 to 150 mmHg, increments of 10 mmHg*
Lower auto-retry
programmable limits:
10 to 100 mmHg, increments of 10 mmHg*
Pulse pressure change
(delta) limits:
30 to 100 mmHg, increments of 10 mmHg*
Deflation rate: True linear deflation according to AHA guidelines.
Programmable for 2, 3, 4, 5, or 6 mmHg per second.
Display: 32 character LCD (16 characters per row) provides a clear
view of data and message prompts.
Sample quantity: Over 250 samples per set of 4 new batteries.
Sample periods: Manual and interval programmable. Intervals of 5, 6, 7, 8, 9,
10, 15, 20, 25, 30, 35, 40, 45, 50, 55, 60, 90, 120 or 240
minutes can be selected/programmed.