3-16 ApexPro™ 2001989-351A
Installation and configuration
2. At the PC, use a communication program such as HyperTerminal to set up the
ComPort connection.
3. Press Enter to get the @> Enter Service Password: prompt.
4. Type password, mms_aps (case-sensitive).
5. Press Enter.
6. At the @> prompt type ssf.
The command is “ssf x” where x identifies which antenna field(s) to enable.
Enter the “ssf x” command as listed below to configure the desired antenna
field(s).
7. Exit the communication program, then disconnect serial cable and PC, or
continue with step 6 in Setup the receiver on page 3-17.
Port: Com1 (comm port on the PC)
Speed: Baud 19200
Parity: No
Stop Bit: 1
Data Bits: 8
Flow Control: Xon Xoff
X Field D Field C Field B Field A
1 Disabled Disabled Disabled Enabled
2 Disabled Disabled Enabled Disabled
3 Disabled Disabled Enabled Enabled
4 Disabled Enabled Disabled Disabled
5 Disabled Enabled Disabled Enabled
6 Disabled Enabled Enabled Disabled
7 Disabled Enabled Enabled Enabled
8 Enabled Disabled Disabled Disabled
9 Enabled Disabled Disabled Enabled
10 Enabled Disabled Enabled Disabled
11 Enabled Disabled Enabled Enabled
12 Enabled Enabled Disabled Disabled
13 Enabled Enabled Disabled Enabled
14 Enabled Enabled Enabled Disabled
15 Enabled Enabled Enabled Enabled