5-8 ApexPro™ 2001989-351A
Troubleshooting
a. Perform Out-of-Band RF Signal Test.
b. Record data in the Clinical Systems Design Tool.
c. If test passes, go to the next test.
d. If test fails:
i. Troubleshoot signal(s) to suspect antenna(s) and add appropriate
filtering.
ii. Verify by repeating the test.
iii. Update documentation.
3. Verify the antenna system meets the requirements of the Unity Signal Gain Test
with Transmitter.
a. Perform the Unity Signal Gain Test with Transmitter.
b. Record data in the Clinical Systems Design Tool.
c. If test passes, go to the next test.
d. If test fails:
i. Verify that the actual installation complies with the system design.
ii. Adjust components as necessary to comply with the design.
iii. Verify by repeating the test.
iv. Update documentation if appropriate.
v. If test passes, go to the next step.
vi. If test still fails, contact GE technical support.
4. Verify the antenna system noise floor.
a. Perform Noise Floor Performance Test.
b. Record data in the Clinical Systems Design Tool.
c. If test fails:
i. Contact GE technical support.
ii. Notify the Field Product Specialist and Director of Service.
d. If test passes, go to the next test.
5. Verify the antenna system meets antenna coverage requirements.
a. Perform Antenna System Coverage Test.
b. Record data in the Clinical Systems Design Tool.
c. If test passes, go to the next test.
d. If test fails:
i. Determine the location(s) of additional antennas needed.