General Ultrasound system requirements
Ultrasound System – Common Service Information 3-13
Direction 5444964-100 English Rev. 5
Electrical Requirements for Docking Carts
Table 3-10:  Docking Cart for LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid| e/Vivid iq Power 
Requirements
PARAMETER AREA LIMITS FREQUENCY
Voltage Range  (Docking 
Cart used for LOGIQ e/
LOGIQ e Vet R7.x.x and 
below, LOGIQ i and Vivid 
e)
100-120V~
220-240V~
500VA
500VA
50/60 Hz
Voltage Range (Docking 
Cart used for LOGIQ e 
R8.x.x or higher)
100-240V~ 350 VA
Vivid iq Docking Cart 100-240V~  190VA ~ 350 VA
Power All applications More than or equal to 750 
VA
Line Frequency All applications 50/60Hz (±2Hz)
Power Transients All applications Less than 25% of 
nominal peak voltage for 
less than 1 millisecond 
for any type of transient, 
including line frequency, 
synchronous, 
asynchronous, or 
aperiodic transients.
Decaying Oscillation All applications Less than 15% of peak 
voltage for less than 1 
millisecond.