GE LOGIQ F SERIES
DIRECTION 5446617-100, REVISION 10 BASIC SERVICE MANUAL
8 - 12 Section 8-5 - Loading Base Image Software
Section 8-5 Loading Base Image Software (cont’d)
8.) System USB memory stick will be loading. Wait for the software installation to complete. The
process is indicated on the screen.
9.) After the applying process completes, the system begins to backup C Dsik.
WARNING
While the software installation procedure is designed to preserve data, select choice
[2] to format disk C only.
WARNING
If selecting [1], please make sure the patient data and images have been backed up
first, then upgrade the system.
Figure 8-24 Loading Process
Figure 8-25 Backing Up