Electrical safety tests
Ultrasound SystemCommon Service Information 4-39
Direction 5444964-100 English Rev. 5
Generic procedure on probe leakage current (continued)
NOTE: Not all test procedures are applicable to all areas of the world.
Reversed Polarity testing content satisfies regions following IEC
62353:2007 and IEC 60601-1:2005.
NOTE: Values in italics font are given as examples only.
Table 4-16: Typical data format for recording probe (source) leakage current
Unit under test___________________________ Date of test:___________
Test Conditions Probe as measured in saline bath
System Power Grounding /PE 4C i12L TS E8C
off closed
off open
on closed
on open
Table 4-17: Typical data format for recording probe (source) leakage current
Unit under test____________________________________ Date of test:_____________
Test Conditions Probe as measured in saline bath
System
Power
Grounding/
PE
Limit
µA
4C i12L TS E8C
off closed 100
off open 500
on closed 100
on open 500
off closed
(reversed
polarity)
100
off open (reversed
polarity)
500
on closed
(reversed
polarity)
100
on open (reversed
polarity)
500

Table of Contents

Other manuals for GE LOGIQ S8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ S8 and is the answer not in the manual?

GE LOGIQ S8 Specifications

General IconGeneral
TypeUltrasound System
Display TypeLCD
3D/4D ImagingYes
ConnectivityDICOM, USB, Ethernet
Display Size19 inches
TransducersLinear, Convex, Phased Array
Doppler ModesColor Doppler, Power Doppler, Spectral Doppler
Operating ModesB-Mode, M-Mode
ApplicationsAbdominal, Cardiac, Obstetrics, Gynecology, Vascular, Musculoskeletal
PortabilityCart-based
Transducer Ports4

Related product manuals